Configure the Modality for Storage
Almost done with the Ultrasound set up. The last task assigned to you was to configure the modality for storage. Let's go ahead and set that up.
This time you were careful. You edited the image storage settings. Configured the AET in all caps and double checked your work. Let's perform a verify below. Also, to get more information, turn on debugger mode and press verify again to ensure there is a successful DICOM association with PACS.
DICOM Service - Image Storage
Service List
| Device | Service Name | AE Title | Port | Host |
|---|---|---|---|---|
| Production Storage | Store | PACSPROD | 104 | 192.168.1.2 |
C-ECHO Debug Logs
Take a look at the logs. You'll see that whoever initiated the DICOM association (connection) is the "Calling Application Entity". The one the receives the request is the "Called Application Entity". It looks like the modality called the server and received a successful E=Echo response. Therefore...
The DICOM verification request (C-ECHO-RQ) was successful!
Looks like we're ready to send some test images